Lack circuit

Created: 27/4/2004
Updated: 6/12/2017
 

Lack circuit

A co-axial modification of the Mapleson A system, designed to facilitate scavenging of expired gas. (Dr A Lack, retired Salisbury anaesthetist).

Circuit

Lack construction



Figure 1: The Lack circuit


A four-way block is attached to the fresh gas outlet (F). This block is connected to an outer reservoir tube (R) attached to the patient (P), an inner exhaust tube (E), a breathing bag (B) and a spring-loaded expiratory valve (V).

Function

Functioning of the Lack circuit
Figure 2: Function of the Lack circuit

The Lack circuit is essentially similar in function to the Magill, except that the expiratory valve is located at the machine-end of the circuit, being connected to the patient adapter by the inner coaxial tube.

Inspiration - The valve closes and the patient inspires fresh gas from the outer reservoir tube.

Expiration - The patient expires into the reservoir tube. Towards the end of expiration, the bag fills and positive pressure opens the valve, allowing expired gas to escape via the inner exhaust tube.

Expiratory pause - Fresh gas washes the expired gas out of the reservoir tube, filling it with fresh gas for the next inspiration.

Controlled ventilation

 As for the Magill circuit.

Advantages

 The location of the valve is more convenient, facilitating intermittent positive pressure ventilation and scavenging of expired gas.

Disadvantages

 In common with other co-axial systems, if the inner tube becomes disconnected or breaks, the entire reservoir tube becomes dead-space. This can be avoided by use of the 'parallel Lack' system, in which the inner and outer tubes are replaced by conventional breathing tubing and a Y-piece: 

Parallel Lack system

Figure 3: 'Parallel Lack' system


ArticleDate:20040427
SiteSection: Article
 
   
    
                                            
  Posting rules

     To view or add comments you must be a registered user and login  




Login Status  

You are not currently logged in.
UK/Ireland Registration
Overseas Registration

  Forgot your password?








 
All rights reserved © 2020. Designed by AnaesthesiaUK.

{Site map} {Site disclaimer} {Privacy Policy} {Terms and conditions}

 Like us on Facebook 

vp